Industrial Technologies

division of BIOLAB 

  

 Cannon Glass Capillary Viscometers & Viscosity Standards


Asphalt Institute Vacuum (AIV) Viscometers

 

 

PDF Datasheet | Request Info


 

 

 

In most glass capillary viscometers, the samples flow under gravity. When liquids are too viscous to flow readily under gravity, vacuum viscometers may be used to measure viscosity (in mPa•s or cP). In these instruments a vacuum is applied to one end of the viscometer to pull the liquid through the capillary into the timing bulb(s). CANNON offers several types of vacuum viscometers, including the Cannon-Manning Vacuum , the Asphalt Institute Vacuum, and the Modified Koppers Vacuum. Like the Cannon-Fenske Opaque viscometer, these are all “reverse flow” viscometers. Vacuum viscometers require a vacuum that is very accurately controlled. The CANNON DVR-1000 or DVR-1500 Digital Vacuum Regulator is an ideal instrument for maintaining constant vacuum.

AIV Calibrated Series
For measurement of viscosity of highly viscous materials such as asphalt cement at 60°C (140°F) according to ASTM D 2171. Applicable to materials with viscosity range of 0.036 to 80 000 poise. Vacuum of 300 mm Hg is applied to small arm with timing bulbs; two bulbs are incorporated to extend viscosity range of viscometer. Similar to the Cannon-Manning Vacuum, but with graduated capillary instead of two timing bulbs.

AIV-RO Calibrated Series
Permanently attached round metal holder, for hole 51mm (2 inches) in diameter. Provided with certificate of calibration and instruction sheet. For sizes 400R and 800R, the "roofing" design is normally used for segment B, below, which extends the viscosity range to 1400000 and 58000000 poise, respectively. Minimum sample volume 3mL. Require liquid bath depth of 180 mm (7.1 inches).

 

Size

Viscosity range, poise

Approx. constant poise/second
at 300mm Hg Bulb B

Approx. constant poise/second
at 300mm Hg Bulb C
Approx. constant poise/second
at 300mm Hg Bulb D
AIV Calibrated Series

 AIV Calibrated RO Series

25    

42 - 800

2

1

0.7

9724-R50

9724-T50

50

180 - 3200

8

4

9724-R55

9724-T55

100 

600 - 12800

32

16

10 

9724-R60

9724-T60

200

2400 - 52000

128

64

40

9724-R65

9724-T65

400R

9600 - 1400000

500

250

160

9724-R70

9724-T70

800R

38000 - 5800000

2000

1000

640

9724-R75

9724-T75

*Viscometers with special constants, added temps, and before/after data are available by special order at an additional cost.