 Cannon Glass Capillary Viscometers & Viscosity Standards


Ubbelohde (UB) Viscometers

 

 

PDF Datasheet | Request Info
 

Ubbelohde viscometers are for use with transparent liquids. Unlike the Cannon-Fenske Routine viscometer, suspended level viscometers possess the same viscometer constant at all temperatures. This property is advantageous when measurements are to be made at a number of different temperatures. Cannon has improved the design of the Ubbelohde viscometer to make a more rugged instrument called the Cannon-Ubbelohde viscometer. Other suspended level viscometers include the BS/IP/SL, BS/IP/SL(S), and BS/IP/MSL viscometers.

Calibrated UB Series
For measurement of kinematic viscosity of transparent Newtonian liquids by suspended level principle as described in ASTM D 445 and D 446, and ISO 3104 and 3105. Similar to Cannon-Ubbelohde but less rugged in design.

UBC Unity Factor Series
Constant within +/- 0.2% of approximate constant.
Provided with certificate of calibration and instruction sheet.
Minimum sample volume 11mL. Required liquid bath depth of 241 mm (9.5 in).

* 300 sec minimum flow time. All other sizes have flow times of 200 to 1000 seconds.

Size

Approx. constant,
mm2/s2(cSt/s)

Kinematic Viscosity Range
mm2/s(cSt)

Uncalibrated
UBU Series
Calibrated
UBC Series
Calibrated UBXC
Unity Factor Series
0* 0.001 0.3 - 1 9721-N50 9721-R50 9721-U50
0C 0.003 0.6 - 3 9721-N53 9721-R53 9721-U53
0B 0.005 1 - 5 9721-N56 9721-R56 9721-U56
1 0.01 2 - 10  9721-N59 9721-R59 9721-U59
1C 0.03 6 - 30 9721-N62 9721-R62 9721-U62
1B 0.05 10 - 50 9721-N65 9721-R65 9721-U65
2 0.1 20 - 100 9721-N68 9721-R68 9721-U68
2C 0.3 60 - 300 9721-N71 9721-R71 9721-U71
2B 0.5 100 - 500 9721-N74 9721-R74 9721-U74
3 1.0 200 - 1000 9721-N77 9721-R77 9721-U77
3C 3.0 600 - 3000 9721-N80 9721-R80 9721-U80
3B 5.0 1000 - 5000 9721-N83 9721-R83 9721-U83
4 10   2000 - 10000 9721-N86 9721-R86 9721-U86
4C    30 6000 - 30000 9721-N89 9721-R89 9721-U89
4B 50    10000 - 50000 9721-N92 9721-R92 9721-U92
5 100  20000 - 100000 9721-N95 9721-R95 9721-U95

*Viscometers with special constants, added temps, and before/after data are available by special order at an additional cost.