Ubbelohde viscometers are for use with
transparent liquids. Unlike the Cannon-Fenske
Routine viscometer, suspended level viscometers possess the same
viscometer constant at all temperatures. This property is advantageous
when measurements are to be made at a number of different temperatures.
Cannon has improved the design of the Ubbelohde viscometer to make a
more rugged instrument called the Cannon-Ubbelohde
viscometer. Other suspended level viscometers include the
BS/IP/SL,
BS/IP/SL(S), and BS/IP/MSL
viscometers.
Calibrated UB Series
For measurement of kinematic viscosity of transparent Newtonian liquids
by suspended level principle as described in ASTM D 445 and D 446, and
ISO 3104 and 3105. Similar to Cannon-Ubbelohde
but less rugged in design.
UBC Unity Factor Series
Constant within +/- 0.2% of approximate constant.
Provided with certificate of calibration and instruction sheet.
Minimum sample volume 11mL. Required liquid bath depth of 241 mm (9.5
in). |
|
* 300 sec
minimum flow time. All other sizes have flow times of 200 to 1000
seconds.
|
Size |
Approx. constant,
mm2/s2(cSt/s) |
Kinematic Viscosity Range
mm2/s(cSt) |
Uncalibrated
UBU Series |
Calibrated
UBC Series |
Calibrated UBXC
Unity Factor Series |
|
0* |
0.001 |
0.3 - 1 |
9721-N50 |
9721-R50 |
9721-U50 |
|
0C |
0.003 |
0.6 - 3 |
9721-N53 |
9721-R53 |
9721-U53 |
|
0B |
0.005 |
1 - 5 |
9721-N56 |
9721-R56 |
9721-U56 |
|
1 |
0.01 |
2 - 10 |
9721-N59 |
9721-R59 |
9721-U59 |
|
1C |
0.03 |
6 - 30 |
9721-N62 |
9721-R62 |
9721-U62 |
|
1B |
0.05 |
10 - 50 |
9721-N65 |
9721-R65 |
9721-U65 |
|
2 |
0.1 |
20 - 100 |
9721-N68 |
9721-R68 |
9721-U68 |
|
2C |
0.3 |
60 - 300 |
9721-N71 |
9721-R71 |
9721-U71 |
|
2B |
0.5 |
100 - 500 |
9721-N74 |
9721-R74 |
9721-U74 |
|
3 |
1.0 |
200 - 1000 |
9721-N77 |
9721-R77 |
9721-U77 |
|
3C |
3.0 |
600 - 3000 |
9721-N80 |
9721-R80 |
9721-U80 |
|
3B |
5.0 |
1000 - 5000 |
9721-N83 |
9721-R83 |
9721-U83 |
|
4 |
10 |
2000 - 10000 |
9721-N86 |
9721-R86 |
9721-U86 |
|
4C |
30 |
6000 - 30000 |
9721-N89 |
9721-R89 |
9721-U89 |
|
4B |
50 |
10000 - 50000 |
9721-N92 |
9721-R92 |
9721-U92 |
|
5 |
100 |
20000 - 100000 |
9721-N95 |
9721-R95 |
9721-U95 |
*Viscometers with special
constants, added temps, and before/after data are available by special
order at an additional cost. |