division of BIOLAB |
|
||||||||||||||||||||||||||||||||||||||||||||||||||||
|

Measurements Group Strain Gauge Instrumentation
Model V/E-40 Strain Gauge Simulator
|
The V/E-40 can also be used in conjunction with a conventional Wheatstone bridge strain indicator to measure arbitrary resistances between 30.00 and 1111.10 ohms, or to eliminate Wheatstone bridge nonlinearity effects when measuring high post-yield strains in quarter-bridge operation. In this mode, the resistance or strain
gauge to be measured is connected as one arm of a Wheatstone bridge, the V/E-40 is used as a decade resistor in an adjacent arm, and the strain measuring instrument as a null detector.
Other applications include use as an investigative tool to troubleshoot faulty strain
gauge installations, or as a precision decade resistor.
Features
5 Decade Selector Switches
Resistance Range: 30.00 to 1111.10 ohms in 0.01 ohm steps
High Precision Vishay Resistors used throughout to ensure Excellent Stability
Accuracy 0.02% of Setting
Simulates Tension and Compression Strain for most widely used Strain Gauge Resistance Values
Simulates a Broad Range of RTD's for Instrumentation Setup and Calibration
Specifications
All specifications nominal or typical at +23 deg C unless noted.
Accuracy:
0.02% of reading.
Maximum Current (To meet accuracy and repeatability specifications.):
120 ohms: 65 mA;
350 ohms: 55 mA;
1000 ohms: 25 mA.
Stability:
+3 ppm/deg C max.
Resistance Range:
30.00 to 1111.10 ohms in 0.01 ohm steps.
Environment:
0 deg to +120 deg F (-18 deg to +49 deg C), up to 70% relative humidity, non-condensing.
Size: 3-7/8 H x 9-1/8 W x 3-1/8 D in (98 x 232 x 89 mm).
Weight: 1.9 pounds (0.85 kg).
|
Copyright © 1997-2001. Vishay Measurements Group, Inc. |